MSE PRO cuvettes for UV-Vis, near-infrared and fluorescence spectroscopy, in optical glass and quartz, with path lengths from 5 mm to 50 mm. Choosing the right cuvette comes down to three things: the wavelength range of your measurement, the path length your sample concentration needs, and the cuvette style your instrument and method call for.
1. Window material: match it to your wavelength
| Material | Usable range | Use it for |
|---|---|---|
| Optical glass | 340 to 2500 nm | Visible-range work such as colorimetric assays, where UV transmission is not needed |
| ES quartz (UV-grade fused silica) | 190 to 2500 nm | UV measurements such as nucleic acids at 260 nm and proteins at 280 nm, plus visible work |
| IR quartz | 250 to 3500 nm | Near-infrared measurements beyond 2500 nm, along with UV and visible |
Optical glass absorbs strongly below about 340 nm, so it gives false readings in the UV. If you are unsure, quartz covers more of the spectrum.
2. Path length: match it to your concentration
Absorbance rises in proportion to path length (the Beer-Lambert law). A 10 mm path is the standard most methods assume. Use a shorter path (5 mm) for concentrated or strongly absorbing samples to keep readings in the linear range, and a longer path (20 to 50 mm) for dilute samples or weakly absorbing liquids where you need more signal.
3. Cuvette style: match it to your instrument and method
- Standard cuvettes have two clear windows for absorbance measurements. Examples: 5 mm optical glass with lid, 10 mm IR quartz with stopper, 40 mm ES quartz with lid.
- Fluorescence cuvettes have four clear windows so emitted light can be collected at 90° to the excitation beam. Examples: 10 mm IR quartz, 20 mm ES quartz, and micro-volume versions in ES quartz and IR quartz.
- Micro cuvettes with white walls have a narrower sample chamber, so they need less sample, with masked side walls that limit stray light. Examples: 5 mm IR quartz, 10 mm optical glass with lid, 10 mm ES quartz with stopper, 20 mm optical glass, 20 mm IR quartz, 30 mm ES quartz.
- Self-masking cuvettes have a blackened or masked body around a reduced chamber, which keeps the beam passing only through the sample. Examples: 10 mm ES quartz, 30 mm ES quartz.
- Cylindrical cuvettes are round cells for instruments with cylindrical holders and for longer path lengths. Examples: 20 mm ES quartz, 50 mm IR quartz.
- Round-bottom cuvettes are easier to clean and drain fully. Examples: 5 mm ES quartz, 40 mm optical glass.
- Absorptiometer oblong cuvettes fit absorptiometers and colorimeters with oblong cell holders. Examples: 5 mm IR quartz, 50 mm IR quartz.
- Liquid sample pools (Lovibond type) are cells for Lovibond-style colorimeters used to grade the color of oils, fuels and other liquids. Examples: 6.5 mm ES quartz, 10 mm IR quartz.
Lids and stoppers
Lids keep dust out and slow evaporation during short measurements. Stoppers seal more tightly, which matters for volatile solvents, longer kinetic runs and samples that must not exchange gas with the air.
Care
Handle cuvettes by the frosted or opaque sides, never the optical windows. Rinse immediately after use with the sample solvent, then with deionized water, and avoid abrasive cleaning that scratches the windows. Match the cuvette to the instrument's holder width and beam height before buying, and use the same cuvette for blank and sample where possible.
Looking for the instrument? See laboratory spectrometers and spectrophotometers.