Aconitase(ACO) Activity Assay Kit

  • $ 39295


Aconitase(ACO) Activity Assay Kit

SKU # E-BC-K656-M
Detection Instrument Microplate reader (230-250 nm, optimum wavelength: 240 nm)
Detection method Colorimetric method


Product Details

Properties

Synonyms ACO
Sample type Animal tissue
Sensitivity 2.32 U/L
Detection range 2.32-60.19 U/L
Detection Method Colorimetric method
Assay type Enzyme Activity
Assay time 25 min
Precision Average inter-assay CV: 9.500% | Average intra-assay CV: 5.900%
Other instruments required Centrifuge
Other reagents required Normal saline (0.9% NaCl) , PBS(0.01 M,pH 7.4)
Storage -20℃
Valid period 12 months


Dilution of Sample

The recommended dilution factor for different samples is as follows (for reference only):

Sample type  Dilution factor
10% Rat liver tissue homogenate 6-10
10% Mouse kidney tissue homogenate 4-6
10% Mouse liver tissue homogenate 6-10
10% Mouse brain tissue homogenate 4-6
10% Rat lung tissue homogenate 6-10
10% Mouse muscle tissue homogenate 2-4

 

Note: The diluent is extraction solution B. For the dilution of other sample types, please do pretest to confirm the dilution factor.

 

Detection Principle

Aconitase catalyzes isocitrate to produce cis-aconitase. Cis-aconitase has a characteristic absorption peak at 240 nm. The activity of this enzyme was calculated by measuring the production rate of cis-aconitase.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Extraction Solution A 50 mL × 1 vial 50 mL × 2 vials -20℃, 12 months
Reagent 2 Extraction Solution B 25 mL × 1 vial 50 mL × 1 vial -20℃, 12 months
Reagent 3 Protease Inhibitor 0.7 mL × 1 vial 1.4 mL × 1 vial -20℃, 12 months,
shading light
Reagent 4 Buffer Solution 15 mL × 1 vial 30 mL × 1 vial -20℃, 12 months
Reagent 5 Substrate 0.15 mL × 2 vials 0.15 mL × 4 vials -20℃, 12 months,
shading light
UV Microplate 96 wells No requirement
Plate Sealer 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three mouse liver tissue samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 7.50 22.60 48.50
%CV 10.0 9.4 9.1


Inter-assay Precision

Three mouse liver tissue samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 7.50 22.60 48.50
%CV 5.3 5.6 6.8


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 103%.


Sample 1 Sample 2 Sample 3
Expected Conc. (U/L) 11.2 28.5 53
Observed Conc. (U/L) 11.2 27.9 56.7
Recovery rate (%) 104 98 107

 

Sensitivity

The analytical sensitivity of the assay is 2.32 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.