Trypsin Activity Assay Kit

  • $ 30695


Trypsin Activity Assay Kit

SKU # E-BC-K843-M
Detection Instrument Microplate reader (400-410 nm, optimum wavelength: 405 nm)
Detection method
Colorimetric method


Product Details

Properties

Sample type Animal tissue
Sensitivity 1.86 U/L
Detection range 1.86-100.65 U/L
Detection method Colorimetric method
Assay type Enzyme Activity
Precision Average inter-assay CV: 7.2-11.2% | Average intra-assay CV: 3.9-5.1%
Other instruments required Incubator
Storage -20℃
Valid period 12 months


Dilution of Sample

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
10% Mouse small intestine tissue homogenate 2-5
10% Mouse large intestine tissue homogenate 2-5
10% Rat intestine tissue homogenate 2-5
10% Rat chyme tissue homogenate 2-5

 

Note: The diluent is normal saline (0.9% NaCl). For the dilution of other sample types, please do pretest to confirm the dilution factor

 

Detection Principle

Trypsin catalyzes the substrate to form p-nitroaniline, which has a certain light absorption value at 400-410 nm wavelength. Since the absorbance of p-NA is proportional to the content, the activity of trypsin can be calculated by measuring the p-NA produced per unit time.

 

Kit Components & Storage

Item Component Size 1(48 T) Size 2(96 T) Storage
Reagent 1 Buffer Solution 50 mL × 1 vial 50 mL × 2 vials -20℃, 12 months
Reagent 2 Substrate Power ×1 vial Power ×2 vials -20℃, 12 months, shading light
Reagent 3 Standard Power ×1 vial Power ×2 vials -20℃, 12 months, shading light
Reagent 4 Diluent 2 mL × 1 vial 4 mL × 1 vial -20℃, 12 months
Microplate 96 wells No requirement
Plate Sealer 2 pieces

 

Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

Technical Data:

Parameter:

Intra-assay Precision

Three rat intestine tissue samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 5.60 26.80 76.40
%CV 5.1 3.9 4.5


Inter-assay Precision

Three rat intestine tissue samples were assayed 20 times in duplicate by three operators to determine precision between assays.

Parameters Sample 1 Sample 2 Sample 3
Mean (U/L) 5.60 26.80 76.40
%CV 11.2 7.2 9.8


Recovery

Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 106%.


Standard 1 Standard 2 Standard 3
Expected Conc. (mmol/L) 0.24 0.5 0.73
Observed Conc. (mmol/L) 0.3 0.5 0.8
Recovery rate(%) 113 100 105

 

Sensitivity

The analytical sensitivity of the assay is 1.86 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

 

Standard Curve

As the OD value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

Concentration (mmol/L) 0 0.2 0.3 0.4 0.6 0.7 0.8 1.0
Average OD 0.367 0.423 0.454 0.484 0.543 0.571 0.595 0.659
Absoluted OD 0 0.056 0.087 0.117 0.176 0.204 0.228 0.292