MSE PRO™ Salt Spray Filter Device

MSE PRO™ Salt Spray Filter Device is mainly used for salt spray testing machine exhaust filter, in the use of salt spray testing machine, the exhaust gas contains sodium chloride, direct discharge outdoor will cause corrosion damage to the metal materials around the exhaust port, in order to avoid the use of equipment caused by the loss, it is very necessary to install salt spray filter device outside the exhaust port.

Specification:

Product name  Salt spray filter device
Model  MA0066232
External size (mm)  About 1000×1000×700 W×H×D
Air inlet size(mm)  Ф90/50
Exhaust outlet size(mm)  Ф90/50
Net weight  About 90kg
Performance
Use environment  Ambient temperature is +5~50℃
Fog discharge capacity  40m3/h
Intake salt concentration  0~10%
Discharge salt concentration ≤0.2%
Effluent treatment  Discharged after mixing and diluting clean water
Material  PVC plate welded
Power supply information
Power information  AC(220±10)V 50Hz single phase with three wires
Power supply condition  The grounding resistance of the protective ground wire is less than 4Ω. TN -s mode power supply or TT; Requires the user to the installation site for air or power switch device configuration corresponding capacity, and this switch must be independent for the use of this equipment.
Power capacity 

AC220C 50HZ total watt 500w

Max current:2.5A(power socket need to be reserved on site:one piece of 10A socket,the distance between the socket and the power connection of the equipment should be less than 2.5 meters)

Electrical control system
Program controller  The digital display programmable timer controls the drainage time. It enables the automatic replacement of sprinkler water supply at regular intervals, which can be set arbitrarily and operates in an infinite loop. The timing units are seconds, minutes, tenths, hours, and ten hours, and can be switched. Each group can have a maximum timing of 999 hours and a minimum timing of 1 second.
Chemical pump The acid and alkali corrosion-resistant chemical pump automatically pumps the water from the water storage tank, boosts its pressure, and sprays it for a circulation process.
Waste gas filtration system
Principle of Waste filtration
Use consumable parts The third-level filter material is an activated carbon filter screen. It
needs to be replaced regularly, and the replacement period is 6 months.