Glycogen Fluorometric Assay Kit

  • $ 85095


Glycogen Fluorometric Assay Kit

SKU # E-BC-F040
Detection Instrument Fluorescence microplate reader (Ex/Em=535 nm/587 nm)
Detection Method Fluorescence method


Product Details

Properties

Sample Type Animal liver, muscle tissue
Sensitivity 0.06 μg/mL
Detection Range 0.06-4.0 μg/mL
Detection Method Fluorescence method
Assay Time 50 min
Precision Inter-assay CV: 6.6% | intra-assay CV: 3.4%
Storage This product can be stored at -20°C for 12 months with shading light.
Valid period 12 months

 

Dilution Of Sample

The recommended dilution factor for different samples is as follows (for reference only):

Sample type Dilution factor
10% Rat liver tissue homogenate 3000-5000
10% Mouse muscle tissue homogenate 10-20

 

The diluent is buffer solution A. For the dilution of other sample types, please do pretest to confirm the dilution factor.

 

Detection Principle

Glycogen produces glucose under the action of starch glycosidase, and glucose is catalyzed by glucose oxidase to produce hydrogen peroxide. In the presence of the peroxidase, hydrogen peroxide be oxidized to produce the fluorescence substrate. The fluorescence intensity at the excitation wavelength of 535 nm and emission wavelength of 587 nm is proportional to the glycogen content.

 

Kit Components & Storage

Item Component Size (96 T) Storage
Reagent 1 Buffer Solution A 30 mL ×1 vial -20℃, 12 months
Reagent 2 Buffer Solution B 8 mL × 1 vial -20℃, 12 months
Reagent 3 Probe 0.24 mL × 1 vial -20℃, 12 months shading light
Reagent 4 Enzyme Reagent A Powder × 1 vial -20℃, 12 months
Reagent 5 Enzyme Reagent B Powder × 1 vial -20℃, 12 months
Reagent 6
0.1 mg/mL Glucogen
Standard Solution
0.5 mL × 1 vial -20℃, 12 months
Black Microplate 96 wells No requirement
Plate Sealer 2 pieces


Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.

 

    Technical Data:

    Parameter:

    Intra-assay Precision

    Three rat liver samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)

    Parameters Sample 1 Sample 2 Sample 3 
    Mean (μg/mL) 0.15 1.00 2.50
    %CV 4.0 3.2 3.0

     

    Inter-assay Precision

    Three rat liver samples were assayed 20 times in duplicate by three operators to determine precision between assays.

    Parameters Sample 1 Sample 2 Sample 3 
    Mean (μg/mL) 0.15 1.00 2.50
    %CV 7.3 6.2 6.3

     

     

    Sensitivity

    The analytical sensitivity of the assay is 0.06 μmol/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.

     

    Standard Curve:

    As the fluorescence value of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique or temperature effects), so the standard curve and data are provided as below for reference only:

    Concentration (μmol/L) 0 0.5 1.0 1.5 2.0 2.5 3.0 4.0
    Fluorescence value 1269 2695 3650 5304 6826 8119 9197 11360
    1257 2599 3519 5209 6885 7861 9271 11325
    Average fluorescence value 1263 2647 3585 5257 6856 7990 9234 11343
    Absoluted fluorescence value 0 1384 2322 3994 5593 6727 7971 10080