Introduction

The Buck Scientific HC-404 Total Petroleum Hydrocarbon (TPH) Analyzer is a fixed-wavelength infrared spectrometer designed for fast, accurate, and reliable analysis of total petroleum hydrocarbons in water, soil, and sludge samples.

Operating at a fixed wavelength of 2924 cm⁻¹ / 2930 cm⁻¹, the HC-404 enables rapid TPH measurement without unnecessary scanning. The system supports commonly used analytical methods, including ASTM D7066-04, EPA Method 418.1, and EPA Method 413.2, making it suitable for environmental testing, industrial wastewater monitoring, laboratory screening, oil site discharge testing, automotive service station waste monitoring, and marine transportation discharge analysis.

With intuitive controls, digital readout options, and a compact footprint, the HC-404 provides a simple and dependable solution for laboratories requiring routine TPH analysis with minimal service requirements.

Key Features

  • Fixed-wavelength infrared spectrometer for rapid TPH analysis
  • Suitable for water, soil, and sludge samples
  • Supports ASTM D7066-04, EPA 418.1, and EPA 413.2 methods
  • Digital readout in % transmittance, absorbance, or concentration mode
  • Compatible with 10, 50, and 100 mm quartz cells for low-level detection
  • Peltier-cooled PbSe detector with excellent sensitivity and stability
  • Compact benchtop design with intuitive controls
  • Made in the USA

Typical Applications

  • Environmental water and soil quality analysis
  • Underground storage tank closure testing
  • Laboratory screening of samples for expected organic content
  • Oil site storage tank washing discharge monitoring
  • Industrial wastewater discharge monitoring
  • Automotive service station waste and water discharge monitoring
  • Marine bilge and ballast discharge testing

Specification

Application Total petroleum hydrocarbon analysis in water, soil, and sludge samples
Supported Methods ASTM D7066-04, EPA 418.1, EPA 413.2
Wavelength Fixed at 2924 cm⁻¹ / 2930 cm⁻¹
Wavelength Equivalent 3.42 microns
Bandwidth 0.04 microns
Source Hot quartz halogen source
Detector Type Peltier-cooled PbSe detector
Signal-to-Noise Ratio Better than 2500:1
Readout Modes % Transmittance, absorbance, concentration
Transmission Range 0–100% T
Absorbance Range 0–2 absorbance units
Concentration Range 0.1–1,000 concentration
Detection Capability To 1 ppm or lower with appropriate quartz cells
Accuracy ±1%
Drift 0.01 absorbance units/hour
Response Time 10 seconds in damping 1; 40 seconds in damping 2
Compatible Cell Path Lengths 10 mm, 50 mm, or 100 mm quartz cells
Calibration Method Simple 3-step calibration
Recorder Output Optional
Power Requirements 115V / 230V AC, 50–60 Hz
Power Consumption 80 W
Dimensions 14" W × 11" D × 10" H
Weight 20 lbs
Country of Origin Made in the USA


Optional Parts

Accessories

BS 21-I-10 Infrasil quartz cuvette rectangular, 10mm pathlength
BS 21-I-10-6 Above cuvette in a package of 6
5708 10mm cuvette holder
BS 34-I-50 Infrasil quartz cuvette cylindrical, 50mm pathlength
5710 50mm cylindrical cuvette holder
BS 34-I-100 Infrasil quartz cuvette cylindrical, 100mm pathlength
5707 100mm cylindrical cuvette holder

Freon Method Accessories

404-11 EPA Method 413.2 / 418.1 TPH (used to make you own standards)

Non-Freon / Hexane Method Accessories

404-13 Vessel Cell for Non-Freon Method
404-14 Calibration Standards in Hexane 50ml. (10, 50, 250ppm)
6105 Holder for Vessel Cell

Spare Parts

404-1032 Drive Belts (5pack)
404-1004 Source lamp (new lamp will not fit into old holder as of serial # 990)
404-1004A Upgrade holder with Lamp, one time purchase for Serial # 989 or earlier