ATTO WSE-3200 Power Station III - MSE Supplies LLC

ATTO WSE-3200 Power Station III

SKU: WSE-3200

  • $ 2,89995


Request a Quote

ATTO WSE-3200 Power Station III

Buy ATTO products from MSE Supplies at the best Value.

Purpose and Application:

  • Slab/mini-slab electrophoresis (polyacrylamide gel electrophoresis)
  • Small disc (isoelectric point) Electrophoresis
  • Agarose gel electrophoresis
  • Semi-dry rotting can be used
    under certain conditions (please contact us for conditions for use in blotting)

Features:

  • ~1000V, ~500mA, ~200W output configurable
  • Programmable output settings Free: 21 files, chained: 10 files
  • Output conditions such as electrophoresis have already been input (preset)
  • Original output conditions can be saved (memory)
  • Large dial for easy configuration
  • Reliable accuracy, stability, and safety features
  • Easy-to-read display Large LED, two-stage display

 

Applicable WSE-3100 WSE-3400 WSE-3500 WSE-3200
Wet (Tank) blotting  ⭕️  ❌  ⭕️  ❌ 
Submerge Agaros gel  ⭕️  ⭕️  🔺  ⭕️ 
Semi-Dry blotting  ⭕️  🔺  ⭕️  🔺 
Plate IEF Smaller than 10 x 10 cm  🔺  ❌  ❌  🔺 
PAGE gel Smaller than 10 x 10 cm  ⭕️  ⭕️  🔺  ⭕️ 
PAGE gel Larger than 10 x 10 cm  ⭕️  ⭕️  ❌  ⭕️ 
Large PAGE (Sequence gel)  🔺  ❌  ❌  🔺 
Faster Semi-Dry blotting  ⭕️  🔺  ⭕️  🔺 
Disc IEF  ⭕️  🔺  ❌  ⭕️ 
Cooling PAGE  ⭕️  ❌  ❌  ⭕️ 
  ⭕️ Applicable  🔺 Restricted by condition  ❌ not applicable   

 

 

MSE Supplies is an authorized distributor of ATTO products.

Technical Data:

Model and name WSE-3200 Power Station III.
output Voltage: 10~1000V (Display 1000=Full) (Max. 200W) Current
: 0~500mA (Display 0~) (Max. 200W) Power
: 0~200W (Display 0~)
establishment Voltage: 10~1000V 1V steps Current
: 1~500mA 1mA steps Power
: 1~200W 1W steps
Output terminals 4 sets of parallel outputs
timer 1~999min, continuous 1min step countdown
Control Functions Constant Voltage, Current, and Power Crossover
Control unit Display: 7seg LED 3 digits x 2 columns
Setting: Large Dial
Alarm Output Electronic buzzer key input (1 time), error occurrence (1 time), time up (5 times)
Output accuracy Voltage: 10~50V or less for set value±1.5V 51V
~150V for set value ±3%+5digits 151V
~1000V ±2%+5digits for set value
Display accuracy Voltage: Within ± (0.2% of RDG+5digits) Current
value: Within ± (0.5% of RDG+5digits) Power
: Within ± (0.5% of RDG+5digits)
Conditional memory Preset: 23 files P01~P23
Free: 21 files U00~U20, 10 files C01-C10 (chain files)
Safety Features Output Short Circuit Detection: Short-Circuit Current (Short) Leak
Detection: Leakage Current to Ground (Leakage) Open
Detection: Load Release (Open)
Arc Detection: Discharge State (ARC)
Input High Voltage Detection: Input Voltage 265V or More
Fan Stop Detection: Fan Failure
Internal temperature detection: equipment internal temperature rise
Measures against power outages After power is restored, it automatically returns to the state before the power outage.
Cooling method Forced air cooling with built-in fan
Usage environment Temperature 0°C~+40°C, Humidity 20~90%RH Non-condensing
Input voltage AC100~240V
frequency 50/60Hz
Power Consumption 300W or less
Safety Standards Compliant with the Electrical Appliance and Material Safety Act
material Case ABS
size 96(W)x325(D)x195(H)mm
mass 2.4kg