Orchid Scientific Diffusion Cell Apparatus without External Temperature Sensor

Order upon Request
SKU: JFDC 06- O30
Regular price $ 9,072.95

Still have questions? Ask our experts!

Orchid Scientific Diffusion Cell Apparatus without External Temperature Sensor

IN-VITRO SKIN PERMEATION STUDIES OF TOPICAL
FORMULATIONS

The Vertical Franz Diffusion Cell is a simple, reproducible test for measuring the in vitro drug release from creams, ointments and gels. The Franz Cell consists of two primary chambers separated by a membrane. The test product is applied to the membrane via
the top chamber- donor compartment. The bottom chamber receptor compartment contains fluid from which samples are taken at regular intervals for analysis. This testing determines the amount of active drug that has permeated the membrane at each time point.

Features:

  • Vibration-free electromagnetic stirrer.
  • External temperature sensor probe to determine temperature of media inside the cell (For Model JFDC07-O30)
  • Data logging for stirring speed and temperature at a sampling interval.
  • Multi-language screen selection.
  • Touch screen display with 20 in-built programs.
  • The same system can be upgraded to multi-station by adding one more stirrer unit.
  • Audio/visual alarms for sampling intervals.
  • Stirring on/off option prior to sampling.
  • Highly accurate stirring speeds and temperature.
  • Heated circulating water bath with water level sensors to avoid damage to the heating element.
  • Double wall Heated circulating water bath ensuring more precise temperature control.
  • Password-protected data collection and report generation software with admin features.
  • Calibration report generation facility in the software.
  • Graphical presentation of data.
  • Provision to add experiment title & comment.
  • Data can be converted to Excel & Pdf files for further analysis.

The Transdermal Diffusion cell apparatus is remarkably simple to operate; the system is supplied with:

  • Six or seven-stage magnetic stirrer with cell holders.
  • Heated circulating water bath with water level sensor.
  • Touch screen control unit for control of speed and temperature.
  • Diffusion Cells & accessories pack.
  • Jacked reservoir beaker for diffusion media.

Specifications:

Specification Model
JFDC 06- O30
(Single Stirrer Unit)
JFDC 06- O30 x 2
(Dual Stirrer Unit)
Stirrer Drive
Stirrer drive 6 stage electromagnetic:
01 (JFDC 06-O30)
6 stage electromagnetic:
02 (JFDC 06-O30 x 2)
Material of construction of enclosure S.S. 304
RPM Indicator Digital (Touch Screen)
Stirring Speed Range & Accuracy 200- 800 rpm (±1.0%)
Heated Circulating Water Bath (WCS02)
Material of construction of enclosure Double wall S.S. 304
Temperature range and accuracy 5°C Above ambient to
60°C (±0.5 °C)
Inter cell temperature variation ±1.0 °C
Inter cell stirring speed variation ±1.0 %
Temperature Indicator Digital (Touch screen)
Certifications CE Compliant
Power requirements 220/230V AC 50Hz,
110/120V AC 50-60Hz*
External temperature sensor NO

 

Ordering Information:

Model Stirring drive Heated
circulating
water bath
Cells Supplied Accessories
supplied
Optional
JFDC
06-O030
6 stage
electromagnetic
stirrer drive: 01
Double wall
Heated
circulating
water bath: 01
unit
-Glass diffusion
cell of 20ml
capacity with
20mm
exposure area:
06Nos
(20ml cells are
by default. If
user wants to
select any
other capacity
of cells, same
can be done
with prior
notification*)
-Cell Clamps:6,
Adequate Silicon
Tubing for
Interconnection.
-Lid for cell
sampling arm: 6
-Stirring bars: 6
-Silicon rings:6
-Dosage
compartment:6
-Support disc:6
-Alignment ring:6
-Luer lock
connectors for
cell
Interconnections:
6 pairs
-IQ, OQ
documents
-Ethernet cable
-Software DVD,
Inlab calibration
certificate for
Diffusion cell and
equipment
-Diffusion
cell of 5ml,
7ml, 12ml
-Precise
sample
withdrawal
glass
syringe
-NABL
accredited
lab
calibration
certificate
for diffusion
cell
-PVT
documentati
on
JFDC 06-
O30 x 2
(Dual
stirrer
unit)
6 stage
electromagnetic
stirrer drive: 02
Double wall
Heated
circulating
water bath:
01 unit
Glass diffusion
cell of 20ml
capacity with
20mm
exposure area:
12 Nos
(20ml cells are
by default. If
user wants to
select any
other capacity
of cells, same
can be done
with prior
notification*)
-Cell Clamps:12
Adequate Silicon
Tubing for
Interconnections
-Lid for cell
sampling arm: 12
-Stirring bars: 12
-Silicon rings:12
-Dosage
compartment:12
-Support disc:12
-Alignment ring:12
-Luer lock
connectors for
cell     Interconnections:
12pairs
-IQ, OQ
documents
-Ethernet cable
-Software DVD,
Inlab calibration
certificate for
Diffusion cell and
equipment

 

 

 

Performance Verification Test (PVT) for Diffusion Cell Apparatus:
(Suitable for Model JFDC 07-O30)
The Vertical Diffusion Cell (VDC) is an in vitro laboratory device for the study of drug release and permeation of semi-solid topical formulations and topical patches. Franz Diffusion Cell is a type of VDC. It is the industry standard for IVRT as well as IVPT studies. However, this Franz diffusion cell needs to pass physical acceptance criteria and performance verification test before performing IVRT/IVPT studies.


Benefits of Performance Verification Test (PVT) a USP requirement:

  • Globally recognized method for Metrology aspects: PVT covers both mechanical and analytical parts of franz diffusion cell apparatus.
  • Ability to uncover the influence of unforeseen system variables in franz diffusion cell apparatus.
  • Ability to compare intra/inter lab variation: PVT is a holistic test and involves system suitability testing by reference standard material (USP 1% Hydrocortisone cream), a validated reference standard cream with pre-established testing characteristics.
  • A necessity for IVRT/IVPT studies

PVT test will cover:

  • Physical Acceptance Criteria according to USP General Chapter <725> and <1724> for parameters like the capacity of cells, the diameter of the orifice, the temperature of receptor medium, the temperature of the membrane surface, and magnetic stirrer speed.
  • Performance Verification Test of VDC Systems using the USP Hydrocortisone Cream for parameters like Cumulative amount of drug (Hydrocortisone) release per square cm and Confidence Interval using HPLC as an analytical method.