Mouse C-P (C-Peptide) CLIA Kit
SKU: E-CL-M0227
Mouse C-P (C-Peptide) CLIA Kit
| Detection Range | 31.25-2000 pg/mL |
| Sensitivity | 18.75 pg/mL |
Product Details
Properties
| Assay type | Sandwich-CLIA |
| Format | 96T |
| Assay time | 3.5h |
| Detection range | 31.25-2000 pg/mL |
| Sensitivity | 18.75 pg/mL |
| Sample type &Sample volume | serum, plasma and other biological fluids; 100μL |
| Specificity | This kit recognizes Mouse C-P in samples. No significant cross-reactivity or interference between Mouse C-P and analogues was observed. |
| Reproducibility | Both intra-CV and inter-CV are < 15%. |
| Application | This CLIA kit applies to the in vitro quantitative determination of Mouse C-P concentrations in serum, plasma and other biological fluids. |
Test Principle
Kit components & Storage
| Item | Specifications | Storage |
|---|---|---|
| Micro CLIA Plate(Dismountable) | 96T: 8 wells ×12 strips 48T: 8 wells ×6 strips |
-20℃, 12 months |
| Reference Standard | 96T: 2 vials 48T: 1 vial |
|
| Concentrated Biotinylated Detection Ab (100×) | 96T: 1 vial, 120 μL 48T: 1 vial, 60 μL |
|
| Concentrated HRP Conjugate (100×) | 96T: 1 vial, 120 μL 48T: 1 vial, 60 μL |
-20℃(Protect from light), 12 months |
| Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8°C, 12 months |
| Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
| HRP Conjugate Diluent | 1 vial, 14 mL | |
| Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
| Substrate Reagent A | 1 vial, 5 mL | 2-8℃(Protect from light) |
| Substrate Reagent B | 1 vial, 5 mL | |
| Plate Sealer | 5 pieces | |
| Manual | 1 copy | |
| Certificate of Analysis | 1 copy |
Technical Data
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse C-P were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse C-P were tested on 3 different plates, 20 replicates in each plate.
| Intra-assay Precision | Inter-assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean (pg/mL) | 93.46 | 276.82 | 787.41 | 86.95 | 253.12 | 821.88 |
| Standard deviation | 9.49 | 26.74 | 47.56 | 8.07 | 17.95 | 90.32 |
| CV (%) | 10.15 | 9.66 | 6.04 | 9.28 | 7.09 | 10.99 |
Recovery
The recovery of Mouse C-P spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
| Sample Type | Range (%) | Average Recovery (%) |
|---|---|---|
| Serum(n=8) | 90-104 | 97 |
| EDTA plasma(n=8) | 93-105 | 99 |
| Cell culture media(n=8) | 88-102 | 94 |
Target Information
| Synonyms | Connecting Peptide |
| Research Area | Signaling transduction,Metabolism |
Assay Procedures
![]() |
1.Add 100 μL of standard or sample to each well. Incubate for 90 min at 37℃. |
![]() |
2.Remove the liquid. |
![]() |
3.Add 100 μL Biotinylated Detection Ab. Incubate 1 hour at 37℃. Aspirate and wash 3 times. |
![]() |
4.Add 100 μL HRP Conjugate. Incubate 30 min at 37℃. Aspirate and wash 5 times. |
![]() |
5.Add 100 μL Substrate Mixture Solution. Incubate for 5 min at 37℃. |
![]() |
6.Fluorescence appeared. Measure the RLU value with the Chemiluminescence immunoassay analyzer. Calculation of the results. |