Monoamine Oxidase (MAO) Activity Assay Kit
Couldn't load pickup availability
Consumables Product
This consumable product is not available for refund or return
Still have questions? Ask our experts!
Monoamine Oxidase (MAO) Activity Assay Kit
| SKU # | E-BC-K008-M |
| Detection Instrument | Microplate reader(345 nm-360 nm, optimum wavelength: 355 nm) |
| Detection Method | Colorimetric method |
Product Details
Properties
| Synonyms | MAO |
| Sample Type | Animal tissue |
| Sensitivity | 16 U/L |
| Detection Range | 16 – 641 U/L |
| Detection Method | Colorimetric method |
| Assay Time | 40 min |
| Precision | Inter-assay CV: 6% | intra-assay CV: 3.3% |
| Storage | This product can be stored at 2~8°C for 12 months with shading light. |
| Valid period | 12 months |
Dilution Of Sample
The recommended dilution factor for different samples is as follows (for reference only):
| Sample type | Dilution factor |
|---|---|
| 10% Mouse liver tissue homogenate | 1 |
| 10% Rat liver tissue homogenate | 1 |
| 10% Rat lung tissue homogenate | 1 |
| 10% Rat kidney tissue homogenate | 1 |
| 10% Mouse brain tissue homogenate | 1 |
The diluent is buffer solution working solution. For the dilution of other sample types, please do pretest to confirm the dilution factor.
Detection Principle
MAO can catalyze 4-dimethylambenzylamine to produce p-dimethylaminobenzaldehyde. p-Dimethylaminobenzaldehyde has a characteristic absorption peak at 355 nm. The activity of MAO can be calculated indirectly by analyzing the production of p-dimethylaminobenzaldehyde.
Kit Components & Storage
| Item | Component | Size 1(48 T) | Size 2(96 T) | Storage |
| Reagent 1 | Extraction Solution A | 30 mL ×1 vial | 60 mL ×1 vial | 2-8°C, 12 months |
| Reagent 2 | Extraction Solution B | 60 mL ×1 vial | 60 mL × 2 vials | 2-8°C, 12 months |
| Reagent 3 | Buffer Solution | 60 mL ×1 vial | 60 mL × 2 vials | 2-8°C, 12 months |
| Reagent 4 | Chromogenic Agent | 3 mL ×1 vial | 5 mL ×1 vial | 2-8°C, 12 months |
| UV Microplate | 96 wells | No requirement | ||
| Plate Sealer | 2 pieces | |||
Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.
Technical Data:
Parameter:
Intra-assay Precision
Three rat liver tissue samples were assayed in replicates of 20 to determine precision within an assay. (CV = Coefficient of Variation)
| Parameters | Sample 1 | Sample 2 | Sample 3 |
|
Mean (U/L) |
25.00 | 240.00 | 350.00 |
| %CV | 3.5 | 3.1 | 3.3 |
Inter-assay Precision
Three rat liver tissue samples were assayed 20 times in duplicate by three operators to determine precision between assays.
| Parameters | Sample 1 | Sample 2 | Sample 3 |
|
Mean (U/L) |
25.00 | 240.00 | 350.00 |
| %CV | 5.7 | 6.2 | 6.1 |
Recovery
Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 100%.
| Sample 1 | Sample 2 | Sample 3 | |
| Expected Conc. (U/L) | 90 | 285 | 460 |
| Observed Conc. (U/L) | 9.27 | 305.0 | 483.0 |
| Recovery rate (%) | 103 | 107 | 105 |
Sensitivity
The analytical sensitivity of the assay is 16 U/L. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.