Malondialdehyde (MDA) Colorimetric Assay Kit (Cell Samples)
Couldn't load pickup availability
Consumables Product
This consumable product is not available for refund or return
Still have questions? Ask our experts!
Malondialdehyde (MDA) Colorimetric Assay Kit (Cell Samples)
| SKU # | E-BC-K028-M |
| Detection Instrument | Microplate reader (530-540 nm, optimum wavelength: 532 nm) |
| Detection Method | Colorimetric method |
Product Details
Properties
| Synonyms | MDA |
| Sample Type | Cells |
| Sensitivity | 0.29 nmol/mL |
| Detection Range | 0.29-100 nmol/mL |
| Detection Method | Colorimetric method |
| Assay type | Quantitative |
| Assay time | 65 min |
| Precision | Average inter-assay CV: 3.500% | Average intra-assay CV: 3.300% |
| Other instruments required | Micropipettor, Centrifuge, Incubator, Water bath, Vortex mixer |
| Other reagents required | Absolute ethanol |
| Storage | 2-8℃ |
| Valid period | 12 months |
Dilution of Sample
It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.29-100 nmol/mL).
| Sample type | Dilution factor |
| A549 cells | 1 |
| HepG2 cells | 1 |
| 293T cells | 1 |
Note: The diluent is reagent 5.
Detection Principle
MDA in the catabolite of lipid peroxide can react with thiobarbituric acid (TBA) and produce red compound, which has a maximum absorption peak at 532 nm.

Kit Components & Storage
| Item | Component | Size (500 Assays) | Storage |
| Reagent 1 | Clarificant | 30 mL × 1 vial | 2-8℃, 12 months |
| Reagent 2 | Acid Reagent | 30 mL × 1 vial | 2-8℃, 12 months |
| Reagent 3 | Chromogenic Agent | 50 mL × 3 vials | 2-8℃, 12 months shading light |
| Reagent 4 | 10 nmol/mL Standard | 5 mL × 1 vial | 2-8℃, 12 months |
| Reagent 5 | Extracting Solution | 30 mL × 1 vial | 2-8℃, 12 months |
Note: The reagents must be stored strictly according to the preservation conditions in the above table. The reagents in different kits cannot be mixed with each other. For a small volume of reagents, please centrifuge before use, so as not to obtain sufficient amount of reagents.
Technical Data:
Parameter:
Intra-assay Precision
Three HepG2 cell samples were assayed in replicates of 20 to determine precision within an assay (CV = Coefficient of Variation).
| Parameters | Sample 1 | Sample 2 | Sample 3 |
| Mean (nmol/mL) | 2.60 | 34.50 | 88.50 |
| %CV | 3.6 | 3.2 | 3.1 |
Inter-assay Precision
Three HepG2 cell samples were assayed 20 times in duplicate by three operators to determine precision between assays.
| Parameters | Sample 1 | Sample 2 | Sample 3 |
| Mean (nmol/mL) | 2.60 | 34.50 | 88.50 |
| %CV | 3.1 | 3.8 | 3.6 |
Recovery
Take three samples of high concentration, middle concentration and low concentration to test the samples of each concentration for 6 times parallelly to get the average recovery rate of 95%.
| Sample 1 | Sample 2 | Sample 3 | |
| Expected Conc. (nmol/mL) | 12 | 49 | 91 |
| Observed Conc. (nmol/mL) | 11.2 | 48.0 | 85.5 |
| Recovery rate (%) | 93 | 98 | 94 |
Sensitivity
The analytical sensitivity of the assay is 0.29 nmol/mL MDA. This was determined by adding two standard deviations to the mean O.D. obtained when the zero standard was assayed 20 times, and calculating the corresponding concentration.