Performance Characteristics
Heating bath circulation thermostats with a stainless-steel bath allow the temperature control of internal and external thermoregulation tasks. Items can be directly thermoregulated by placing them in a thermostat bath and the control of externally connected applications is also possible. The units are suitable for a number of typical temperature control tasks such as sample temperature control, materials testing, quality control, etc. The devices are fitted with overtemperature and low-level protection in line with Safety Class III/FL according to DIN 12876. A powerful pressure/suction pump ensures optimal circulation and temperature homogeneity in the thermostat bath and also allows temperature control of externally connected applications.
Functions overview
- Pilot ONE controller with Plug & Play technology
- Precise temperature control
- 5.7“touchscreen and comfortable menu navigation
- Overtemperature and low-level protection
- Safety class III/FL (DIN 12876)
- Powerful pressure/suction pump
- Programmer with calendar and clock functions
- 2-point calibration
- Pt100 sensor connection
Technical Data
|
General & Temperature Control |
|
|
Temperature range |
28...300 °C |
|
with cooling unit |
-20...300 °C |
|
with water cooling |
20...300 °C |
|
Temperature stability |
±0,02 K |
|
Controller |
Pilot ONE |
|
Safety class |
III / FL |
|
Protection class |
IP20 |
|
Dimensions (W x D x H) |
210 x 335 x 392 mm |
|
Weight, net |
14 kg |
|
Heating / cooling capacity |
|
|
Heating capacity |
2,2 - 3 kW |
|
Circulation pump |
|
|
Delivery capacity pressure pump max. |
25 l/min; 0,7 bar |
|
Delivery capacity suction pump max. |
18,5 l/min; 0,4 bar |
|
Pump connection |
M16x1 AG male |
|
Max. permissible viscosity |
50 mm²/s |
|
Pump speed adjustable |
Yes |
|
Pump characteristics |
Pressure and suction pump, variable speed |
|
Connections |
|
|
External sensor connection |
Pt100 |
|
Interface digital |
Ethernet, USB (Host u. Device), RS232 |
|
Operating data |
|
|
Permissible ambient temperature |
5...40 °C |
|
Bath opening WxD/BD |
130 x 100 / 155 |
|
Bath volume |
5 l |
|
Filling volume min. |
3,2 l |
|
Power supply |
200-240V 1~/2~ 50/60Hz |
|
Current consumption max. |
14 A |
|
Fuse |
16 A |
Datasheet
Operating Manuals
Heat-up Curve
Pump Curve